Operation

E-mail Print

Operation:

  1. Open the case
  2. fill the container with 500ml calibration liquid
  3. connect one power supply with the case (pump)
  4. connect the other power supply with the heating element
  5. switch on the heating element with the button onoff
  6. connect the tube with pressure reducer to the pump and heating element
  7. connect the tube without pressure reducer to the heating element and your breathalyser
  8. now wait until the LCD displays a temperature of 34°C and the green LED (READY) flashes

 

CA 0403 Bedienung

 

Calibration:

After the calibration facility is set up and the READY LED flashes, you can begin with the calibration of your breathalyser.

  1. Select “Calibration” in the menu of your breathalyser
    (more details are in the manual of the breathalyser)
  2. Connect the breathalyser with the calibration facility
  3. Start the calibration by pressing the red button of the pump. For about 10 seconds air flows into the breathalyser and at the same time the breathalyser makes a noise. If there is no noise heard the pressure should be raised with the help of the pressure reducer